DIELECTRIMETRO
A DIELECTRIMETER IS A COMBINATION OF 2 EQUIPMENTS DEDICATED FOR 2 FUNCTIONS
- a megohmmeter, made to measure insulation resistance, high resistances
- a hipot tester to make a high voltage test (also called dielectric strength test, hipot test, hi-pot test, breakdown test or flash test)
The dielectrimeter can do automatically and successively those 2 above tests, as hipot then megohm, megohm then hipot, or megohm then hipot then megohm again. This last test is the ideal one, which ensures that the high voltage test has not been damaging anything on the device under test.
APPLICATION
SPECIFICATION OVERVIEW
Dielectrimeter (insulation test + hipot test) selector guide
| Sefelec dielecrimeter range |
Hipot tester / Dielectric strength tester AC voltage
|
Hipot tester / Dielectric strength tester
DC voltage
|
Hipot tester / Dielectric strength tester
Current |
Megohmmeter
Resistance
|
Megohmmeter
Voltage |
Interfaz |
Software |
| Dielectrimetro DXS50 Nuevo |
5kV |
No |
10mA AC |
50kΩ - 200GΩ
2TΩ in option
|
10...500V DC
(1V steps)
1000V DC
in option
|
RS232 included
GPIB, PLC, Ethernet in option
|
SXS Pro |
| Dielectrimetro DXS56 Nuevo |
5kV |
6kV |
10mA AC
4mA DC
|
50kΩ - 200GΩ
2TΩ in option
|
10...500V DC
(1V steps)
1000V DC
in option |
RS232 included
GPIB, PLC, Ethernet in option
|
SXS Pro |
| Dielectrimetro DXS500 Nuevo |
5kV |
No |
110mA AC |
50kΩ - 200GΩ
2TΩ in option
|
10...500V DC
(1V steps)
1000V DC
in option |
RS232 included
GPIB, PLC, Ethernet in option
|
SXS Pro |
| Dielectrimetro DXS506 Nuevo |
5kV |
6kV |
110mA AC
110mA DC
|
50kΩ - 200GΩ
2TΩ in option
|
10...500V DC
(1V steps)
1000V DC
in option |
RS232 included
GPIB, PLC, Ethernet in option
|
SXS Pro |
Dielectrimetro
PREMIER 2725A |
5kV |
No |
40mA AC
|
up to 2GΩ |
50/100/500/
1000V DC |
RS232 in option |
No |
Dielectrimetro
PREMIER 2735A |
5kV |
6kV |
40mA AC
10mA DC
|
up to 2GΩ |
50/100/500/
1000V DC |
RS232 in option |
No |
| Dielectrimetro MPC47P |
6kV |
6kV |
5mA AC 5mA DC
|
500kΩ - 2TΩ |
45/100/250/
500V DC |
GPIB, PLC in option |
No |
APPLICATION
MANUFACTURER, LABORATORY, HIGH SPEED PRODUCTION, MAINTENANCE, IN ORDER TO COMPLY WITH THE LOW VOLTAGE DIRECTIVE

•cable manufacturer (measures made on reel, on cable samples, … > shielding against wires)
•transformers
•components (capacitors, resistances, sensors,…)
•aeronautic, aerospace sub assemblies (satelites panels, …)
•connectors, switches, activators, ...
•test laboratories
•car industry (splices, electric engines, pumps, …)
•satellite receivers, decoders, television receivers, ...
•solar panels (test in between current output and the frame)
SPECIFICATION OVERVIEW
FROM LABORATORY TO PRODUCTION TESTING, OUR RANGE COVERS THE FOLLOWING SPECIFICATIONS
•resistance measurement up to 2 TΩ
•insulation test with DC Voltage from 1V up to 1000 VDC (1V steps)
•hipot test with our ”delta I” exclusive detection type for non destructive testing, up to 5 kVAC, 6 kVDC
•1single output connexion for 2 tests done in automatic testing
•possible destructive hipot testing (for investigation and laboratory use) and hipot at operating temperature
•GPIB, RS232, Ethernet, 0-10V, PLC available
•SXS Pro software for quality purposes, traceability and professional reporting
|